GE Camera Protocol

CEA SCAN ACQUISITION AND PROCESSING
FOR THE GE SYSTEM

ACQUISITION

 

ENERGY SETTING: Tc99m - 140 Kev with a 20% Window.

PLANAR:

OR

SPECT: (Always acquire the Pelvic SPECT before the Abdominal SPECT.)

PROCESSING

PLANAR DISPLAY:

CINE RAW DATA DISPLAY:

SPECT PROCESSING

USE: RECON

**ATTENUATION CORRECTION COMMAND = ADAT:
(Use only for the SPECT of the liver, if necessary.) Occasionally the liver will display a dark rim particularly on the lateral edge of the right lobe. If there is a question of a lesion at the edge of the liver, Attention Correction may help differentiate between a reconstruction artifact and disease. Process the SPECT of the liver with and without attenuation correction and compare the images.

Use Image Triangulation to Evaluate the study.
TRIANGULATION COMMAND = TRIF

Remember to vary the intensity to bring out any subtle structures.

SPECT DISPLAY

SPECT DISPLAY COMMAND = DISP

HELPFUL TIPS:

DISCLAIMER:

This procedure was not developed by, or is not meant as an endorsement by GE of CEA-Scan.  These suggested parameters are based on the experience of various users with this type of equipment and are not a substitute for the physician's or technologist's professional experience or knowledge. Users are ultimately responsible for determining if this procedure is usable with their equipment and standard practices. It is assumed that Quality Control of the imaging equipment is performed according to the manufacturer's recommendations. Please refer to the full product prescribing information and consult your equipment manufacturer's updated user's manual prior to the first use of CEA-Scan.

 